     BLOOD PRESSURE MONITORING                                                                                     Hypertension


     Cuffless blood pressure measurement: Where do we actually stand?
     Mukkamala R, et al. Hypertension. 2025 Jun;82(6):957-970.

     Cuffless BP measurement represents a popular and important   market, recent data regarding pulse wave analysis (PWA) and
     area of biomedical technology, driven by the clinical need for   pulse arrival time (PAT), progress toward validation standards,
     more convenient and comprehensive BP evaluation. Global    and ultimately where the field stands. A central objective is
     hypertension management demands regular monitoring,        to address whether PWA and PAT have provided meaningful
     especially since the most recent guidelines recommend      advances or whether skepticism regarding their reported
     targeting an SBP of about 120 mmHg, and the importance of   accuracy is warranted.
     monitoring BP during sleep for a complete profile. However,
     conventional cuff devices are often inaccessible to all    Current research is dominated by PWA and PAT technology,
     potential users and can provide misleading readings in clinical   which typically involves measuring pulsatile waveforms through
     settings. Cuffless technology has the potential to overcome   photoplethysmography (light) or tonometry (force) and applying
     these limitations, and there has been an increase in related   machine learning algorithms. While many studies report positive
     publications and regulatory-cleared devices. With so much   results, these are often undermined by methodological problems
     growing evidence and technological advancement, why do     like data leakage (where the variable to be predicted is included
     almost all major BP device manufacturers still not offer cuffless   or leaked into the data), the need for frequent individual user
     BP devices?                                                calibration, and small, homogenous study populations. Real-
                                                                world challenges, such as motion artifacts, skin tone differences,
     Despite the rise in interest, the field remains fraught with   vascular variability, and environmental conditions, further
     controversy. The purpose of this review is to provide an   complicate accuracy. Collectively, these factors overestimate
     essential update on the intense activity and ongoing debate   performance and make existing validation protocols, which
     surrounding cuffless BP technologies over the last few years.   were designed for cuff-based devices, inappropriate or even
     It summarises the clinical need, the latest devices on the   misleading for cuffless technologies.

                                 Popular PWA with or without PAT for cuffless BP measurement